small incinerators


The incinerator shall be equipped with water injection nozzle connection. If the temperature in the primary

chamber should rise too high (normal temperature between 300 and 800?C) during the operating time, the automatic water

injection shall be activated to decease the temperature.

 

  1. The incinerator shall have the following features and facilities:-

 

a-    Fuel oil grade No. 2 (diesel oil) fired burners 220V, 50 Hz, single phase 4 wire electricity supply electric

motors insulated to Class “F” suitable for operation at 50?C ambient temperature.

 

b-    The incinerator to be fully assembled wired and logic function tested in the factory, prior to dispatch.

 

c-    One high flue gas temperature indicator/alarm to provide an audio/visual warning in the event of high flue gas

temperatures occurring in two stages:-

 

d-    First stage audible alarm, second stage incinerator shutdown, with one voltage free contact for connection to a

remote alarm system.

 

e-    Flexible wire control linkages to bring under fire air, overfire air and burner slide plate control to front of

incinerator.

 

f-    One fire protection system, comprising fusible links over burners, deadweight fire valve, electrically

interlocked with incinerator controls.

 

g-    Special feeding door with thermal lock to protect the operator from accident.

 

h-    One set of cleaning tools.

 

i-     Two sets of comprehensive written instructions.

 

  1. The incinerator shall consist of a low temperature carbonizing changer (primary chamber) in which the waste

components are converted into gas by heat which is produced through incomplete combustion as the result of a deficiency

of air, and a gas reactor (thermal reactor) in which the gas is burned with a controlled excess of air.

 

  1. The incinerator shall be equipped with water injection nozzle connection. If the temperature in the primary

chamber should rise too high (normal temperature between 300 and 800?C) during the operating time, the automatic water

injection shall be activated to decease the temperature.

 

  1. The incinerator shall have the following features and facilities:-

 

–      Fuel oil grade No. 2 (diesel oil) fired burners 220V, 50 Hz, single phase 4 wire electricity supply electric

motors insulated to Class “F” suitable for operation at 50?C ambient temperature.


incinerator manufacturers

1. SCOPE OF WORK The work specified in the contract will include.
1. Supply, install and commission an incinerator as provided in the technical specifications
2. Provision for adherence to OSHA and other statutory requirement relating to Health and Safety of workers during execution of the works covered in this tender.
3. Assessment and preparation and construction of a standard incinerator shed with one waste storage room ( see the design )

4. Provision of operation and maintenance manuals and drawings complete with maintenance schedule and instructions.

5. Fuel oil storage tank- Mild steel1500 liters capacity

6. Onsite training for operation and maintenance staff.

IMPORTANT
incinerator manufacturers To enable the Employer undertake conclusive evaluation of the tender, tenderers are requested to submit with their offers the detailed specifications, drawings and catalogues with commissioning and maintenance instructions for the Incinerator they intend to supply, install and commission. Tenderers who fail to comply with this requirement fully shall have their tenders rejected. 2. TECHNICAL SPECIFICATIONS OF THE REQUIRED INCINERATOR

2.1 General Description

The plant must have four distinct sections that demonstrate three principles of turbulence, residence time and temperature retention. The regulated sections may include but not limited to:
1. Overall plant layout
2. Feed chamber/charging
3. Primary combustion chamber
4. Particulate scrubbers
5. Acid gas scrubbers.
6. The stack/chimney.
7. Minimum capacity of primary chamber = 0.18 cubic metres
8. Burn rate (rate of waste incineration) = a least 20 kg per hour
9. Should have a full width counterbalance loading door allowing easy access to the main chamber for loading and removing ash.
Supply, Installation and Commissioning of An Incinerator in Eastern Hydros [ Page 28

2.2 Feeding and Charging
Controlled hygienic mechanical feeding that does not negatively influence the air temperature in the primary and secondary chambers of the incinerator shall be used.
2.3 Primary Combustion Chamber.

a) This chamber should be equipped with burners that use low sulphur fuel.
b) The chamber should ensure a minimum exit temperature of 850 degrees centigrade.

2.4 Secondary Combustion Chamber (After burner)

a) Should be fitted with burners that use low sulphur fuel.
b) Ensure secondary air supply is controlled.
c) Ensure residence time is not less than two seconds.
d) Ensure that the gas temperature as measured against the inside wall in the secondary chamber and not in the flame zone is not less than 1100 degrees centigrade.
e) Ensure the emitted oxygen content of the emitted gases is not less than 11%
f) Ensure both primary and secondary combustion temperatures are maintained until all waste is completely combusted.

2.5 Particulate Removers

A mechanical particulate collector must be incorporated after secondary combustion chamber for removal of particulate pollutants entrained in the flue gas stream. The particulate collectors may include any of the following or a combination thereof:
a) Cyclone separator
b) Electrostatic precipitators
c) Fabric filters
2.6 Chimney/Stack

l) The chimney should have a minimum height of 10 metres above ground level and clear the highest point of the building by not less than 3 metres for all roofs. The topography and height of adjacent buildings within 50 metres radius should be taken into account.
m) The chimney should be visible to the operator from the feeding area.
n) The minimum exit velocity should be 10 metres per second and at least twice the surrounding wind speed (Efflux velocity = wind speed x 2) whichever is higher to ensure no down washing of exiting gases.
o) The point for measurement of emissions shall be provided.

2.7 Instrumentation

a) Instrument for determining the inside wall temperature and not burner flame temperature must be provided for both primary and secondary chambers.
b) An audible and visible alarm must be installed to warn the operator when the secondary chamber temperature drops to below the required temperature.
c) A carbon monoxide and oxygen meter/recorder should be provided

d) A smoke density meter/recorder should be provided.
e) A solid particulate meter/recorder should be provided.
f) Any other instrument that may be considered necessary.

2.9 Emission Limits

a) Combustion efficiency (C.E) shall be at least 99.00% where C.E = %Carbon dioxide x 100 / [% Carbon dioxide + carbon monoxide]
b) The temperature of the primary chamber shall be 800 +/- 50 degrees centigrade
c) The secondary chamber gas residence time shall be at least 1 second at 1050 +/- 50 degrees centigrade with 3% oxygen in the stack gas.
d) Opacity of the smoke must not exceed 20% viewed from 50 metres with naked eyes.
e) All the emission to the air other than steam or water vapour must be odourless and free of mist, fume and droplets.
f) A 99.99% destruction and removal efficiency (DRE) for each principal organic hazardous constituent (POHC) in the waste feed where DRE [ ( Win – Wout) /Win ] x 100 where Win = mass feed rate of the POHC in the waste stream fed to incinerator and Wout = mass emission rate of POHC in the stack prior to the release to the atmosphere.
g) The average dioxin and furan concentration in the emission should not exceed 80mg/m3 total dioxins and furans if measured for a period of 6 to 16 hours.

– Manual technical monitoring is shipped with the hardware with all explanations of failures and resolutions
Capacity/island : 4000 kg/day ; 25000 kg/day ; 31000 kg/day ; 48000 kg/day

Burning temperature – over 12000C

Two chambers – minimum

Emissions – dioxin and heavy metaly – better then Kyoto protocol

Emission – monitoring system

Waste unloiding – directly from waste track to incinerator

Disposables Incinerator

INCINERATOR (WASTE INCINERATOR)
Designed to burn disposables that can and should be destroyed on-site. Theses wastes include infectious and

contaminated “red bag.” Surgical dressings, plastic test devices and other wastes
Fast, complete, efficient waste disposal
Dual chamber combustion, Chambers insulated and lined with high temperature refractory. Programmable digital

temperature controls, temperature indicators and charge recorder. Modulating control for fuel and air lowers

fuel consumption.
Minimum installation and start up time
Aluminized steel jacket lined with refractory and firebrick. Monitors recorders and other accessory equipment

should be available.
Charging system:
•    Vertical charge door on primary chamber.
•    Pneumatic or hydraulic ram charging system.
•    Auto shut down of the door after charging.
•    Primary and secondary chamber temperature displayed and recorded (optional)
Charging rate:
•    Up to 100kgs per hour of waste or red bag waste rated at 2200 BTU’s per kg.
Fuel:-     Natural Gas
Capacity (Cubic Mtr):    1.65M3 or more
Ave Capacity:    100kg per hour
Secondary Chamber Capacity (CubicMtr) :3.88 or more
Length (mm):    1880 or better
Width (mm):    1280 or better
Height Incl. Flue (mm):    7700 or better
Ash  Door Opening (mm):    475×900 or better
Min. Operating temperature:    90 degree centigrade
Max, Operating Temperature:    1350 degree centigrade
Residency time in Second Chamber:    2 sec
Temperature Monitoring: Both for primary and secondary chambers.
Loading SystemManual/semi-automatic
Waste destruction efficiency:> 90% by weight
Filter: Ceramic or other equivalent filters in order to reduce the induction of hazardous air pollutants,

particulates, co,dioxin/furane into the atmosphere.
De ashing system: Manual/semi-automatic.
The System must be quoted complete with:-
•    Plumbing
•    Electrical panels
•    Civil works
•    Waste collection Trolleys

Finish/ Paint:-
High Quality heat resistant, two layers high-grade epoxy Paint coating.
Installed Power (Electric Voltage): 440V/ 50 Hz
Waste Management Equipment must have ;
Mobile loading trolley:    06 No’s – For transportation of hospital waste from wards to incinerator site/ room,

Opening from the top.
WASTE BINS MOBILE:    06 No’s-Made of Plastic, in blue color with a cover flap.
Plastic Bags:    2000 no’s each in Red and Yellow color.
Heavy Duty Gloves Pairs: For Waste handling staff-06 Pairs
Long Shoes Pairs:    For waste handling – 06 Pairs.
Make:  UK, EEC, USA.

MEDICAL WASTE INCINERATOR (50 C 60kg/hr)


Medical Waste Incinerator, 50 to 60 Kg/hr
Support Health Sector Support Project
Point of Installation (Hospitals) Moi Voi, Makindu, Maragua, Eldama Ravine and Isiolo District Hospitals
1.    General Description
Supply, delivery, installation and commissioning of a medical waste incinerator suitable for disposal of Medical, General and Pathological waste in a safe and clean environment.  The unit shall consist of two chambers and operate on the principal of controlled air and temperature. The unit shall consist a particulate remover (scrubbers) as stipulated in waste management regulations, 2006 (Legal notice NO. 121 of 29th September, 2006).  The Unit shall be fully automatic and controlled by an automatic electronic controlled system except loading system which shall be manual.  The unit shall be capable of incinerating between 50 to 60 kg of solid medical waste per hour.  It shall be constructed from mild or aluminized steel lined with refractory material.
Main unit
Application For incineration, general and pathological
Capacity   50 C 60 kg/h burn rate
Type Two  combustion chambers type; primary  and Secondary, controlled/forced combustion air type with a flue gas emission scrubbing unit
Operating time                Minimum 8 hours daily
Operating temperature     From 850 0C to 1200 0C, Automatic controlled
Residual Ash                    5 to 10%
    3.2     Primary Chamber
Construction Constructed from heavy duty mild or aluminized  steel Or
equal and approved equivalent
Insulation material            Refractory material lining similar or equal to calcium 
Silicate and hot face combination of heavy duty brickwork
Internal Construction        Fixed hearth type complete with gratings, concave bottom
and charging door, lined with refractory material
Charging Door                   Suitable for manual loading of wastes and with smooth 
Dear seal equivalent of Ceramic seals with hinges.
Door Lock                          Automatic, Electric type
Ash removal door    Provided, for removing resultant bottom ash leftovers                              from the Primary chamber
Gratings   Provided
Loading Manual loading of waste
Primary Burner                        Fully automatic, with fuel, temperature and speed controls with ignition system, flame detector, Air fan complete with safety features, flame failure,  Diesel fired fuel injector type and Flange mounted                                                                                                    
Blower Provided. For supplying excess combustion air through the distribution system with speed control system
Temperature Minimum exit 850 0C
Observation port To be provided with protective glass type
Construction   Constructed from heavy duty mild or aluminized steel or equal and approved equivalent
Insulation Refractory material lining
Combustion Temperatures        Above 850 0C, controlled electronically
Gas residue or retention Time       > 2 second at minimum 850 0C
Secondary Burner                    Provided, Diesel fired, fully automatic, with fuel, temperature and speed controls, With ignition system,                                                                  Flame detector, Air fan, Complete with safety features, flame failure Diesel fired fuel injector type.                                                                  Flange mounted
Ejector Provided, Venturi type, for cooling the flue gases
Combustion Air Fan Provided for supplying combustion and creating a negative drift and turbulences
Temperature Maximum  1600 0C


Large Capability Pet Dog Cremation HICLOVER Version A1500 with E-Power Automobile. Door


Large Capacity Animal Pet Cremation HICLOVER Model A1500 with E-Power Auto. Door

Version

A1500

Inner Measurements

150 x 100 x 96cm

Suit for Animal

Pet cat – – yes

– Dog – yes

Pig – small

Cow – – tiny

Main Product Listing

Main Burning Chamber

Additional Burning Chamber

Mix-Combustion Chamber

Smoke Filter Chamber

Burner Control Case

Chimney:10.0 Meters

*( upgrade to Stainless-steel, cost-free price)

Italy oil/gas heater: 03 sets

Oil Tank (if oil fuel): 500Liters

Design for Pet Dogs Cremation Business

Movable system & & Auto. Door & Fire/Flame Monitoring Opening

Independent Placement for Tiny Pet Dog





















HICLOVER – Medical Environmental 


 
Waste Incinerators
Medical Waste Incinerator
Pet Animal Cremation
Solid Waste Incinerator

Tel:  +86-25-8461 0201   
Mobile: +86-13813931455(whatsapp/wechat)
Website: www.hiclover.com  
Email: [email protected]
Email: [email protected]  
Nanjing Clover Medical Technology Co.,Ltd.

 

2020-06-28


Pre-assembled containerized kind Burner


Pre-assembled, containerized type Incinerator

Design

CA100

Picture

Waste Treatment

Medical Waste Incineration Treatment

Feed Capacity

Average 100-200 kgs per feed

Burning Price

Ordinary 100-120 kgs per hr

Burning Time per Feed

Standard 1-2 hr

Control Mode

PLC( Programmable Reasoning Controller Burner)

Voltage

220V

Power

1.0 Kw

Fuel

Diesel oil

Burner

Italy Heater

Feed Setting

Handbook

Temperature Level Screen

Digital Display

Gas usage (Oil)

Typical 20.4 Kgs/Hour

Inner Measurements

100x80x70cm (Primary Burning Chamber)

External Capacities

230 x 130 x 155cm (main body)

Waste burning chamber

560Liters

Blog Post Burning Chamber

280Liters

Oil Storage Tank Capability

200 Litres

Door Opening

70 x 50cm

Chimney

5.0 M

Gross Weight

4500kgs

Chamber Material

Firebrick, Refractory Concrete

Max. Warm Value

166,000 Kcal/Hr.

Procedure Technical Specifications

Solid Chamber temperature

8000C -10000 C

Gas Chamber temperature level

10000C -12000 C

Chamber Anti-Rate

17500C

Residency time

2.0 Sec.

Burning performance

> 98%

Waste Lower Calorific Power

3000Kcal

Pre-assembled, containerized type do not included chimney installation,containerized  mobile type.
 Burning rate: 100 – 120kg / hour
 Minimum operating temperature: 800-1000 degree
  (Primary combustion)
Maximum operating temperature: 1000-1200 degree  (afterburner chamber)
Secondary Burning capacity: Compulsory.
Density insulation: 50mm high
Refractory casting: minimum 114mm
Heavy duty refracting lining to withstand 1750 degree
Average burn out time: maximum 1-2 hours
Fuel type: fuel oil (diesel)
Fuel consumption in L/hr: average 20.4 liter/Hr.
Residency time in secondary chamber: 2 secs.
Residue ash post 100kg Max 3.8 kg
Installation period Maximum 7days.
State: Anambra Sites:
General Hospital Onitsha
General Hospital Enugwu Ukwu
General Hospital Ekwulobia
Orumba General Hospital
II. State: Benue Sites:
North East Senatorial zonal waste collection site (zone A): EPI Office Katsina-Ala LGA
North West Senatorial zonal waste collection site (zone B): EPI Office Gboko LGA
Benue South zonal waste collection site (zone C): EPI Office Otukpo LGA
Benue State Capital: Central Medical Store, Makurdi
III. State: Kogi Sites:
University Teaching Hospital, Anyigba, Dekina LGA
General Hospital Obangede, Okehi LGA
General Hospital Mopa-Muro LGA
IV. State: Ebonyi Sites:
General Hospital Ezzango, Ebonyi North Zone
 General Hospital Onueke, Ebonyi Central Zone
External L= 3235, W=1150, 1950/3825,Internal L=1600, W=675, H=700;Chamber volume(m3)=0.76;Weight(appox tonne)=3.6,max load capacity (kg) 300, Nominal burn rate uk(kg/hr.)=<50,Thermal capacity(kw) =390, power supply 50/60Hz =220-250v,Door Apertune = 67×550; Hot earth principle = double pass, Fuel type Diesel; Control Panel = Al, Insulation board 25 mm super wool, insulation fire brick =grade E23, Fire brick(Alumina) =42.5%
incinerator capable of a load of upto 50kg/hr.
Twin chamber and should be diesel fuel fired complete with automatic temperature controller upto 900` Celsius.
incinerator (2 equipment) 500kg/hr for face mask disposables.
Automatic smoke free incinerator of capacity 150-200kg/hr for industry. The waste cab be comprises of following types:-
         i.            Wet waste.
       ii.            Plastic wrappers,
      iii.            Solid waste(wood, papers, cartoon box and others).
V. State: Imo Sites:
 Nwabosi Health Center, Nwabosi settlement Isiekenesi ward 2, Ideato South LGA
 Comprehensive Health center, Umuezeala Owerre ward, Ehime Mbano LGA
 Nnenasa Health Center, Umugba settlement, Umueze ward Njaba LGA
General Hospital Owerri Umuguma settlement, New Owerri ward 1
VI. State: Oyo Sites:
State cold store Jericho, South-West LGA, Ibadan
Oyo East Local Government Secretariat, Oyo East LGA
Adebayo Alata Primary Health Care Center (PHC), Ogbomosho South LGA
Itesiwaju NPI Office, Out Itesiwaju LGA
Saki West NPI Office, Saki West LGA
Ibarapa Central NPI Office, Ibarapa Central LGA
VII. State: Kwara Sites:
Ifelodun LGA
Ilorin West LGA
VIII. State: Enugu Sites:
Nsukka District Hospital, Nsukka Town, Nsukka LGA
Agbani District Hospital, Agbani Town, Nkanu West LGA
Udi District Hospital, Udi Town, Udi LGA
IX. State: Enugu MSites:
Odukpani LGA
Obubra LGA
Boki LGA


HICLOVER – Medical Environmental 


 
Waste Incinerators
Medical Waste Incinerator
Pet Animal Cremation
Solid Waste Incinerator

Tel:  +86-25-8461 0201   
Mobile: +86-13813931455(whatsapp/wechat)
Website: www.hiclover.com  
Email: [email protected]
Email: [email protected]  
Nanjing Clover Medical Technology Co.,Ltd.

 

2020-06-30


Small a Animal Crematory Equipment Bargains


Basic Info.

Model NO.:small incinerators for sell
Pullution Sources:Solid Waste Processing
Processing Methods:Combustion
Export Markets:Global

Product Description

How you can discover reasonable incinerator to your regional website?  Incinerator isn’t a cheap equipment, generally the little scale model's cost around $5, 000usd per unit and large scale model's cost more than $50, 000usd per unit. So, before purchase one unit incinerator, you have to make clear following items:  A, burn what? Incinerator is a incineration equipment, function is flame waste. So what sort of waste you would like to treatment?  B, little or large? This according to the waste output daily or the waste volume per feed period, like animal dimensions.  Incinerator use gas or oil fuel, generally, gas fuel is less expensive than petroleum and insenerator double room, insinerator plc, insinerator sampah, installation on high-temperature pyrolysis burning off medical and biowaste china, the smoke cleaner than petroleum fuel model. However, some local website don't have gas distribution system.  Along with other technology items, like power(220v, 380v), waste feed doorway type, incinerator construction, etc. 

A Model animal crematory burning rate from 20-400kgs per hour. Heavy duty combustion room and match for large animal cremation.  Residency time 2.0 sec. In post combustion chamber. Virtually smoke and smell free. 

Smoke Filter Chamber with filter brick, Virtually smoke and smell free after burningoff.

Italy original gas or oil burner, quality dependable.

User guider, complete installation and performance document for customer.

Items/Model A1500 A2400 A4000 A5500
Burning Rate *animal 90 kgs/Hr. 150 kgs/Hr. 200 kgs/Hr. 400 kgs/Hr.
Feed Capacity *animal 120 kgs 200 kgs 300 kgs 500 kgs
Equipment Weight 7000 kgs 8300 kgs 13000 kgs 16500 kgs
Picture  
Primary Chamber (Liters) 1500 2400 4000 5500
Secondary Chamber (Liters) 500 1200 1500 1500
External Dimensions (cm) 240x170x380 260x220x420 320x220x460 360 x 220 x 475
Internal Dimensions (cm) 150x100x96 220x110x100 257x147x108 300 x 147 x 125
Oil Tank(Liters) 500 500 500 500
Door Opening (cm) 59 x 81 90 x 110 108 x 128 108 x 128
Chimney (M) 10 10 14 14
Chimney Type Stainless Steel Stainless Steel Stainless Steel Stainless Steel
Secondary Chamber  YES YES YES YES
Mix-Combustion Chamber YES YES YES YES
Smoke Filter Chamber YES YES YES YES
Combustion Fuel Oil/Gas Oil/Gas Oil/Gas Oil/Gas
Residency Time 2.0 Sec. 2.0 Sec. 2.0 Sec. 2.0 Sec.
Temperature Monitoring YES YES YES YES

waste incinerator manufacturers china


Capacity    Not less than 500Kg/hr.
Operation    24hr-7days/week continuous operation.
Combustion chamber    2 combustion chamber.
Temp.    First combustion chamber 800-1000 C.
Secandry combustion chamber 1000-1200 C, with the possibility to be upgraded to 1400 C.
Combustion Efficiency    99.99%
Retention time    Of the flow gases not less than 2 Sec in the hot portion.
Excess Oxygen concentration    Of primary combustion chamber 3-6% max.
Refractory    The chambers should be lined with high thermal insulating refractory bricks with suitable thickness and with highly temp. resistance. Not less than 1400 C for the first combustion chamber, and not less than 1600 C for the secandry combustion chamber, and at the same time prevent the Temp. of the outside body of the chamber not higher than 70 C.
Doors    Should be lined with the same bricks, and the edges lined with special suitable insulating material to prevent any gas escaping, and the doors should be kept under negative pressure during the operation.
Air Emission Control System    With high destruction and removal efficiency (DRE). Not less than 99.9999 as per KEPA.
Wet System    If the system equiped with a CEMS washing the flow gas with water, then it should be manufactured from suitable anti corrossion materials, and equiped with pH meter to measure and neutralize the water. It should be able to use the water in closed recycled circuit many times before treating/ discharging into the sewage system in compliance with KEPARS.
Chemical dosing    It should be equiped with suitable autochemical dosing system to control the pH for the water of the wet scrubber system including sensors.
Chimney    It should be suitable with the capacity and the technical specifications, with height not less than 12m from ground level. With easy opennable sampling point equal about 3 inch, and the optimum sampling locations is at least 8D downstream direction and 2D upstream from any flow disturbancies.
Sampling point    Should be equiped with a platform located down the sampling point by a distance not less than 1 m, and equiped with suitable stairs from ground to the port with comfortable slope.
Ash Removal    It should be auto ash removal system.
Fuel    It should be dual system, such as diesel and natural gas.
Feeding System    Should be auto feeding system into primary chamber.
Control Room    Should be equiped with a control panel shows Temp. of the combustion burners, gas monitoring, digital Temp. indecation for the primary/secondary chambers, digital Temp. indication for the gases…etc. it should be also Auto operation, burnures on/off auto lamps switch, auto start/stop for air emission control system, auto gas continuous monitoring system reading and print out of gas emission parameters…etc.
Cooling System    The primary chamber should be provided with auto cooling systemthrough nozzles to permit direct cooling if the Temp. will be heigher than 1100 C.
Inter-lock system    The incinerator should be equiped with inter-lock system to prevent feeding the waste to the primary chamber if the combustion Temp. less than 500 C.
Flue treatment and cleaning    To consist a combined system (wet/dry), using water, lime, active carbon and ceramic filters.
Quenching system/Heat exchanger    Should be made and manufactured from anti corrosion and acid attack materials such as stainless steel 316 or equivalent.
Noise Level    Not more than 85 dBA during normal operation.
Inaddition    The desigh should insure that responsive fail-safe control systems are used.
The incinerator should be installed inside closed suitable building with good aeration and ventilation system. The building has facilities for off-loading of wastes from transport vehicles. Auto cleaning system for trucks and waste containers, and cool storage area…etc.
The quotation should included list of original spare parts for 7 years.
The quotation should include supervising, starting up and training of the operators and ministry of health engineers inside/outside Kuwait.
Detailed drawings, catalogs should be included.